Features:
Using import semiconductor laser tube and import semiconductor
photosensitive tube,assure sensitivity and response speed.
Using import micropump,assure
samping volume steady,reduce
noise.
The key component-light scattering sensor obviously eliminate
intracorporal stray light,improve SNR, test result more precise and
reliable.The light scattering sensor got Chinese patent.
Confidence meansurement of this instrument,According
to GB/T6167-2007 Standard,realize
auto estimate clean class,creatively to auto estimate ≤9 location points
clean class.
1000 records stored,free to check previous data.User can optional PC
data analysis sofeware,realize data managerment.
User can chose humiture sensor(optional accessories).
Samping volume display as digit,more intuitive and precise.
“Measurement Number”function,realize
multipoint managerment.
Lithium battery,more
safety,continuous
working 6 hours.
Meets JJF-1190-2008 particulate counter standard
Self-purification time≤10min
Deflection of measurement≤±10%
Error of particulate size 0.5um 5um≤±30%
Error of concentration 0.5um≤±30%
Samping volume and error:2.83L/min
≤±5%
Range of size channel:0.3,0.5,1,3,5,10um,6
channels,0.3um
SNR≥2.5:1
Max concentration:≤300,000/L
Power DC7.2V,1800mAh.
Performance parameter:
Particle Size |
0.3, 0.5,1,3,5,10μm |
Light Source |
Laser diode(using hour>30000 hours) |
Flow |
2.83 L /min(0.1cfm) |
Counting Rate |
50±10%@0.3μm, 100±10%@0.5μm |
Deflection of Measurement |
≤±10% |
Samping Time |
User set (1~99min) |
Samping Delay |
User set (1~99min) |
Samping Rate |
1~99times or continuous |
Alarm Class |
Class10~
Class300,000 class(FED209E Standard)or 2~
9class (ISO14464-1 standard) |
Test Type |
Singleness/Repeat/Continuous/Timing/Average |
Data Storage |
1000 records |
Error Indicator |
Exceed storage,laser died,exceed adjusted flow(>±5%),battery
died |
Communication Interface |
USB、RS232 |
Max Communication Speed |
12MBPS |
Charging Time |
3hours,first
time 12hours |
Self-purification |
≤10min |
Concentration limit |
"+/-5% at 2000000 particles/ft3 (71000 Particle/L) |
Display Type |
LCD big screen |
Print Type |
External printer,Epson ribbon |
Sampling Location |
1000 group |
Counting Type |
Cumulative counting,region |
Power Source |
Chargeable Ni-MH battery(6V/2500mAH)or
AC adapter(input 100~240V) |
Battery working hour |
Continuous working > 6 hours |
Overall Size |
125*43*230mm |
Weight |
0.98kg |
Environmental Conditions |
Working Environment: 15~35℃,
10-85%RH;Storage Environment:-10~50℃,
10-85%RH |
Standard Accessories |
AC
adapter,battery charge,reset filter,isokinetic sampling
probe,tripod,portable protection case,operation
instruction,adjust report,printer |
Optional Accessories |
External humiture sensor,Data
analysis software optical disk and dataline |
Test Standard |
JIS-B-9921(1997), ASTM-F649-01, ASTM-F328-98 (NIST traceable)
JJG547-1988 GB/T6167-2007 JJF1190-2008 |
UCL Estimate Sandard |
ISO14644 FS209E |
|