Rilevatore di difetti ad ultrasuoni MFD510B
RANGE
0~ 9999mm (at steel velocity); range selectable in fixed steps
or continuously variable.
PULSER
Spike excitation with low, middle and high choices of the pulse
energy.
Pulse Repetition Rate: manually adjustable from 10 to 1000 Hz.
Pulse width: Adjustable in a certain range to match different
probes.
Damping: 100©, 200©, 400© selectable to meet different
resolution and sensitivity need.
Probe work mode: Single element, dual element and through
transmission;
Receiver
real-time sampling at 160mhz high speed enough to record the
defect information.
Rectification: positive half wave, negative halfwave, full wave,
and rf
Db step: 0db, 0.1 db, 2db, 6db step value as well as auto-gain
mode
Alarm
alarm with sound and light,.
Mermory
Total 100 configuration channels store all instrument operating
parameters plus dac/avg curve; stored configuration data can be
easily previewed and recalled for quick, repeatable instrument
setup. Total 1000 datasets store all instrument operating
parameters plus a-scan. All the configuration channels and
datasets can be transferred to pc via usb port.
functions
peak hold:
Automatically searching the peak wave inside the gate and hold
it on the display.
Equivalent diameter calculation: find out the peak echo and
calculate its equivalent diameter.
Continuous record: record the display continuously and save it
to the memory inside the instrument.
Defect localization: localize the defect position, including the
distance, the depth and its plane projection distance.
Defect sizing: calculate the defect size
Defect evaluation: evaluate the defect by echo envelope.
Dac: distance amplitude correction
Avg: distance gain size curve function
Crack measure: measure and calculate the crack depth
B-scan: display the cross-section of the test block.
real-time clock
real time clock for tracking the time.
Communication
usb2.0 high-speed communication port
specifications
range: (0 ~ 9999) mm
bandwidth: (0.5 ~ 15) mhz
material velocity: (1000 ~ 9999) m / s
dynamic range: ≥ 32db
vertical linear error: ≤ 3%
horizontal linear error: ≤ 0.2%
resolution:> 40db (5p14)
sensitivity leavings: 60db (flat-bottomed deep hole 200mmф2)
rejection: (0 to 80)% linear
noise level: ≤ 10%
power supply: dc 9v; lithium batteries work for 8 hours or more
ambient temperature: (-20 ~ 50) ℃
relative humidity: (20 ~ 95)% rh
overall dimensions: 263 × 170 × 61 (mm)
Mfd510b standard configuration
No. |
Item |
Quantity |
1 |
Main body |
1 |
2 |
Straight beam probe |
1 |
3 |
Angle probe |
1 |
4 |
Machine-probe cable (q9-q9) |
1 |
5 |
Battery module |
1 |
6 |
Power adapter (charger) |
1 |
7 |
Support pillar |
1 |
8 |
Manual |
1 |
9 |
Instrument case |
1 |
10 |
Datapro software |
1 |
11 |
Usb communication cable |